Use of oil‐free mechanical pumps with leak detectors
作者:
M. Hablanian,
期刊:
Journal of Vacuum Science&Technology A: Vacuum, Surfaces, and Films
(AIP Available online 1991)
卷期:
Volume 9,
issue 3
页码: 2039-2042
ISSN:0734-2101
年代: 1991
DOI:10.1116/1.577451
出版商: American Vacuum Society
关键词: LEAK DETECTORS;HELIUM;VACUUM PUMPS;TURBOMOLECULAR PUMPS;PERFORMANCE;USES;GAS ANALYSIS;MASS SPECTROSCOPY;MASS SPECTROMETERS;ULTRAHIGH VACUUM
数据来源: AIP
摘要:
With the growing demand for contamination‐free leak detectors, the replacement of the oil‐sealed mechanical pumps has become an important issue. Despite their higher total ultimate pressure (0.010 Torr), the oil‐free pumps retain the pumping speed for helium at pressures less than 1×10−7Torr. This is sufficient in most applications using a turbomolecular pump as the high vacuum pump. Experimental results obtained with a Contra‐Flow leak detector with both mechanical pumps replaced by oil‐free pumps are presented. System stability and helium background levels permit leak rate measurements near 2×10−10atm cc/s.