SIMPLIFYING NEUTRON MOISTURE METER CALIBRATION
作者:
F. NAKAYAMA,
R. REGINATO,
期刊:
Soil Science
(OVID Available online 1982)
卷期:
Volume 133,
issue 1
页码: 48-52
ISSN:0038-075X
年代: 1982
出版商: OVID
数据来源: OVID
摘要:
We describe several methods for calibrating different neutron moisture meters. The simplest was found to be that of calibrating one “master” probe and comparing other probes with this under similar field conditions over a range of soil water contents. If several probes are available initially, plastic absorbers of various thicknesses can be used as secondary standards to establish a regression equation for the different probe. From the slope and intercept values obtained, a reference regression equation can be developed that can be used to calibrate other probes in the laboratory.