首页   按字顺浏览 期刊浏览 卷期浏览 A comparative study of plateletpheresis using Baxter Autopheresis C and Haemonetics PCS...
A comparative study of plateletpheresis using Baxter Autopheresis C and Haemonetics PCS Plus

 

作者: J. D. Walton,   E. A. Caffrey,   J. P. Allain,  

 

期刊: Transfusion Medicine  (WILEY Available online 1994)
卷期: Volume 4, issue 1  

页码: 57-61

 

ISSN:0958-7578

 

年代: 1994

 

DOI:10.1111/j.1365-3148.1994.tb00244.x

 

出版商: Blackwell Publishing Ltd

 

关键词: cost effectiveness;donor platelet count;platelet yield;plateletpheresis;quality assurance;specification

 

数据来源: WILEY

 

摘要:

SUMMARY.This study compared plateletpheresis on the Haemonetics PCS Plus (PCS Plus) and the Baxter Autopheresis C (Auto C) using the same 100 selected donors. The number of packs meeting UK BTS/NIBSC specification (>2.2 times 1011platelets per pack) was achieved by 99% of PCS Plus and 82% of Auto C procedures. The positive correlation found between donor precount and final platelet yield was better for the PCS Plus. Both machines met U.K. specification for white‐cell contamination but this was significantly greater for the Auto C. Plasma yields were similar.As a result of this study we chose to use the PCS Plus for routine plateletpheresis in our unit. This has enabled us not only to comply with UK BTS/NIBSC specifications for apheresis platelets easily and cost effectively but also to meet our own higher specification (2.75 times 1011platelets per pack) using existing staff and without extending the working da

 