首页   按字顺浏览 期刊浏览 卷期浏览 Comparison of Nonisotopic and Radioimmunoassay Techniques for DigoxinA Study Based on E...
Comparison of Nonisotopic and Radioimmunoassay Techniques for DigoxinA Study Based on External Quality Assurance Measurements

 

作者: J. Wilson,   L. Tsanaclis,   J. Williams,   J. Tedstone,   A. Richens,  

 

期刊: Therapeutic Drug Monitoring  (OVID Available online 1989)
卷期: Volume 11, issue 4  

页码: 477-479

 

ISSN:0163-4356

 

年代: 1989

 

出版商: OVID

 

关键词: Digoxin;External quality assurance;Accuracy;Precision.

 

数据来源: OVID

 

摘要:

Summary:A comparison of the precision and accuracy of three nonisotopic and of radioimmunoassay (RIA) techniques for digoxin was made using data from 141 samples of spiked human serum assayed between 1984 and 1987 by members of the Heathcontrol external quality assurance scheme. There were significant differences between techniques in the number of observations rejected as outliers greater than 3 SD from the sample mean, the percentages of rejected measurements being Abbott TDX 1.4, RIA 3.3, Ames TDA 4.2, and Syva EMIT 6.3. Ames TDA and Syva EMIT were shown to have a significantly lower precision than the other techniques in measurements of digoxin concentrations <1 nmol/L. A significant 4% negative bias was observed for Ames TDA measurements of digoxin concentrations between 1 and 2.6 nmol/L. A concentration‐related bias was demonstrated for Abbott TDX measurements that varied from 14% overestimates at concentrations <1 nmol/L to 6.7% underestimates of digoxin at concentrations >2.6 nmol/L.

 

点击下载:  PDF (2354KB)



返 回