首页   按字顺浏览 期刊浏览 卷期浏览 Investigation of basic imaging properties in digital radiography. 6. MTFs of II–TV digi...
Investigation of basic imaging properties in digital radiography. 6. MTFs of II–TV digital imaging systems

 

作者: Hiroshi Fujita,   Kunio Doi,   Maryellen Lissak Giger,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 12, issue 6  

页码: 713-720

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.595653

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

We devised a new, simple technique for measuring the modulation transfer function (MTF) of a digital imaging system by using an image of an angulated slit. With this technique, the “presampling” analog MTF, which includes the geometric unsharpness, the detector unsharpness, and the unsharpness of the sampling aperture, can be measured even beyond the Nyquist frequency. A single‐frame image of a slightly angulated slit was employed in order to obtain Fourier transforms of line spread functions at different alignments. The presampling MTF was determined by averaging the two Fourier transforms which we obtained from two extreme alignments (center and shifted) of the slit relative to the sampling coordinate. The presampling MTFs of our digital subtraction angiographic system were determined in two orthogonal directions for three different image‐intensifier modes.

 

点击下载:  PDF (688KB)



返 回