首页   按字顺浏览 期刊浏览 卷期浏览 Determination of Urea in Undiluted Blood Samples by Flow Injection Analysis Using Optos...
Determination of Urea in Undiluted Blood Samples by Flow Injection Analysis Using Optosensin

 

作者: BoA. Petersson,   HelleB. Andersen,   EloH. Hansen,  

 

期刊: Analytical Letters  (Taylor Available online 1987)
卷期: Volume 20, issue 12  

页码: 1977-1994

 

ISSN:0003-2719

 

年代: 1987

 

DOI:10.1080/00032718708078039

 

出版商: Taylor & Francis Group

 

关键词: FIA;urea;ammonia;whole blood;optosensing;enzymatic analysis;gas diffusion;immobilized enzymes

 

数据来源: Taylor

 

摘要:

A miniaturized Flow Injection system for the assay of urea in undiluted whole blood is described. Based on the optical determination of ammonia, the system incorporates a flow cell combining gas diffusion and optosensing, the separating barrier between the donor and accepting streams consisting of a sandwich of a hydrophobic gas permeable membrane and a hydrophilic membrane between which is contained a gel of covalently immobilized urease. The sample urea content is quantified by the colour change of an acid-base indicator contained within the acceptor stream. Within the physiological range the measurement is not affected by variations in the pH value, the buffering capacity or the hematocrit level of the sample solution, nor is it, due to the geometric construction of the flow cell, prone to interferences owing to the colour of the sample solution itself. The system is stable for a t least one week of continuous use.

 

点击下载:  PDF (613KB)



返 回