The calibration and use of a conical hot film anemometer probe in recirculating water flow

 

作者: Michael H. Stenhouse,   Robert L. Stoy,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1974)
卷期: Volume 45, issue 3  

页码: 368-370

 

ISSN:0034-6748

 

年代: 1974

 

DOI:10.1063/1.1686631

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A calibration for directional sensitivity of a conical hot film anemometer probe is described. It is usually assumed that the conical probe is insensitive to the velocity vector direction but, in fact, this is not so because deviations of ±60% of the mean velocity can occur at low water speeds. In this study, the probe is aligned at right angles to the velocity vector but the trend of the results applies to probes aligned along the velocity vector. An application of the conical probe to velocity measurements in a vortical water flow for speeds from 0.06 to 3 m/sec is discussed.

 