A CRITICAL STUDY AND IMPROVEMENT OF THE CHEESE CURD TORSIOMETER2
作者:
A. GERVAIS,
D. VERMEIRE,
期刊:
Journal of Texture Studies
(WILEY Available online 1983)
卷期:
Volume 14,
issue 1
页码: 31-45
ISSN:0022-4901
年代: 1983
DOI:10.1111/j.1745-4603.1983.tb00332.x
出版商: Blackwell Publishing Ltd
数据来源: WILEY
摘要:
The Scott Blair cheese torsiometer designed in the late 60's was rendered automatic by combining an angular transducer with the stress measuring system. A linear transducer was used to verify that the device had sinusoidal movement. The absolute error of the improved system was 3 times lower than that of the original.Gelification kinetics were determined with the automated version. We report an example in which experimental points could be fitted to the Scott Blair model as well as to the Douillard equation (the former is biphasic with an inflection point, and the latter describes a simple exponential curve). In both cases, the standard error of estimation was smaller than or equal to the absolute error of the torsiometer.Even the automated version cannot measure the phase difference between stress and strain with the required precision. Thus, neither the viscosity or elasticity modulus can be determined.