Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system

 

作者: Carey E. Floyd,   Joseph Y. Lo,   Harrell G. Chotas,   Carl E. Ravin,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 18, issue 3  

页码: 408-413

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.596687

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

X‐ray scatter fractions measured with two detectors are compared: a photostimulable phosphor system (PSP) and a conventional film–screen technique. For both detection methods, a beam‐stop technique was used to estimate the scatter fraction in polystyrene phantoms. These scatter fraction measurements are compared to previously reported film‐based measurements. Scatter fractions obtained with the PSP were in good agreement both with measurements using film as well as with most previously reported measurements. For the PSP measurements, repeatability was better than 1%. It was found that the PSP provides a precise x‐ray detector for quantitative scatter measurement in digital radiography.

 