首页   按字顺浏览 期刊浏览 卷期浏览 Transition radiation for the diagnosis of low‐energy electron beams
Transition radiation for the diagnosis of low‐energy electron beams

 

作者: A. H. Mahan,   Alan Gallagher,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1976)
卷期: Volume 47, issue 1  

页码: 81-83

 

ISSN:0034-6748

 

年代: 1976

 

DOI:10.1063/1.1134497

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Electron bombardment of metallic silver is known to produce ’’transition radiation,’’ predominantly at 300–450 nm wavelengths. In reported measurements at 15–500 eV impact energies, transition radiation was used to diagnose the shape and time dependence of an rf modulated electron beam. It is suggested that this is a superior target, as opposed to phosphors or gases normally used, to determine instrumental time response functions and electron beam imaging.

 