Horseradish Peroxidase Enzyme Electrodes for Nadh and Nadph
作者:
Dieter Kirstein,
Florian Schubert,
Frieder Scheller,
期刊:
Analytical Letters
(Taylor Available online 1982)
卷期:
Volume 15,
issue 18
页码: 1479-1491
ISSN:0003-2719
年代: 1982
DOI:10.1080/00032718208065139
出版商: Taylor & Francis Group
关键词: Peroxidase oxidatic reaction;enzyme electrode;NADH;NADPH
数据来源: Taylor
摘要:
The aerobic oxidation of reduced pyridine phosphonucleotides catalyzed by horseradish peroxidase (E.C.1.11.1.7) (HRP) in combination with Mn2+-ions, phenolics and redox dyes was used to perform NADH and NADPH determinations with an amperometric HRP enzyme electrode. Optimal operational conditions have been elaborated for such a sensor and a linear relationship between derivative current (dI/dt) and NADH-concentration was obtained between 2.10−5and at least 2.10−4mol/l. The coefficient of variation was equal to or smaller than 4 %. The time for one measurement was less than 10 minutes including equilibration of the electrode.