Automation of a popular monochromator
作者:
D. C. Look,
J. W. Farmer,
R. N. Ely,
期刊:
Review of Scientific Instruments
(AIP Available online 1980)
卷期:
Volume 51,
issue 7
页码: 968-971
ISSN:0034-6748
年代: 1980
DOI:10.1063/1.1136356
出版商: AIP
数据来源: AIP
摘要:
A wavelength‐scan and intensity‐control system for a widely‐used, high‐intensity monochromator is described. The wavelength scan is bidirectional and variable speed, and may be controlled manually or by TTL logic from a computer. The intensity control is effected by use of a programmable dc power supply and D–A converter. Various filters are described which allow an intensity of up to 2×1014photons/cm2s to be achieved in a 1 cm×3 cm area over a wavelength range 0.76–2.50 &mgr;m, at 0.07 &mgr;m bandwidth, with a single grating. (A lesser intensity is available to 3.4 &mgr;m.) This wide range is made possible by the use of second‐order light from 0.76–1.00 &mgr;m. Photoconductivity data in GaAs:Cr and InP:Fe are presented, as an example.
点击下载:
PDF
(257KB)
返 回