Evaluation of a Linear Self-Scanned Photodiode Array Detector for Direct Detection of 6-8 Kev X-Rays

 

作者: Z. N-C Ha,   S. M. Gruner,  

 

期刊: Instrumentation Science & Technology  (Taylor Available online 1989)
卷期: Volume 18, issue 3-4  

页码: 197-212

 

ISSN:1073-9149

 

年代: 1989

 

DOI:10.1080/10739148908543707

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

The performance of a commercially available 1-dimensional x-ray detector based on a directly bombarded self-scanned photo-diode array has been evaluated for x-rays in the 6-8 KeV range. The detector consists of a linear array of 1024 pixels, each approximately 25 μm wide by 2000 μm tall. Modifications to allow cooling to -55°C, which is required for long x-ray exposures, are described. The signal yield is 0.63 ADC counts/Fe55x-ray, resulting in a detective quantum efficiency greater than 0.5 for doses over a range of 10 to 104x-rays per pixel per scan when operated at -55°C. The resolution is about 3 pixels (75 μm FWHM) and the response with respect to dose is linear up to 1.5 x 104x-rays integrated per pixel per scan. There is some non-linearity at doses below about 1000 x-rays/pixel/scan. The detector has very little geometric distortion. The uncorrected uniformity of response of the individual pixels (field flatness) is good to about 2%. However, the variation in responsiveness is stable and a field flatness calibration can be readily used to correct the uniformity of response.

 