首页   按字顺浏览 期刊浏览 卷期浏览 Design and Performance of a High Momentum, High Luminosity Spectrometer for Particle An...
Design and Performance of a High Momentum, High Luminosity Spectrometer for Particle Analysis

 

作者: J. V. Allaby,   D. M. Ritson,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1965)
卷期: Volume 36, issue 5  

页码: 607-610

 

ISSN:0034-6748

 

年代: 1965

 

DOI:10.1063/1.1719644

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A 90° bendn=0 magnetic spectrometer has been constructed which uses curved field boundaries to achieve second order correction of the focal properties and to make the image plane normal to the emerging particles. The spectrometer has a 112 cm radius of curvature and a maximum momentum capability of 725 MeV/c. The window frame yoke, with sloping edge profiles to minimize edge saturation, enables the focal properties to remain invariant up to 21 kG. The solid angle acceptance is 8×10−3sr with 0.1% resolution and the maximum momentum acceptance is ±5%.

 