首页   按字顺浏览 期刊浏览 卷期浏览 Diminution of Sensitive Time of Spark Chambers
Diminution of Sensitive Time of Spark Chambers

 

作者: Bruce Cork,   Daniel R. George,   T. N. Rangaswamy,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1967)
卷期: Volume 38, issue 2  

页码: 221-223

 

ISSN:0034-6748

 

年代: 1967

 

DOI:10.1063/1.1771359

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The sensitive time of a thin gap spark chamber immersed in a magnetic field was measured. The magnetic field was normal to the electric field of the chamber, and various amounts of Freon 12 (CCl2F2) were used to decrease the sensitive time. A concentration of 100 ppm Freon 12 was sufficient to make the sensitive time less than ½ &mgr;sec, with an efficiency of greater than 70% for delays of less than 300 nsec.

 