Flow Injection Determination of Hydrogen Peroxide Using a Carbon Paste Electrode Modified with a Manganese Dioxide Film
作者:
Klemens Schachl,
Hailemichael Alemu,
Kurt Kalcher,
Jitka Jezkova,
Ivan Svancara,
Karel Vytras,
期刊:
Analytical Letters
(Taylor Available online 1997)
卷期:
Volume 30,
issue 15
页码: 2655-2673
ISSN:0003-2719
年代: 1997
DOI:10.1080/00032719708001812
出版商: Taylor & Francis Group
关键词: hydrogen peroxide;manganese dioxide;carbon paste electrode;amperometry;flow injection analysis
数据来源: Taylor
摘要:
A manganese dioxide film modified carbon paste electrode was developed for use as an amperometric sensor for the determination of hydrogen peroxide (H2O2) in ammoniacal aqueous solutions. The electrode showed a stable response towards H2O2after electrochemical activation. Effects of flow rate, operating potential, concentration, injection volume and interferences were investigated. A linear response towards H2O2from 5 μg.l−1to 450 mg.l−1and a detection limit (3 signal-to-noise ratio) of 4.7 μg.l−1was found. The method was employed for the determination of H2O2in rain water samples.