A manganese dioxide film modified carbon paste electrode was developed for use as an amperometric sensor for the determination of hydrogen peroxide (H2O2) in ammoniacal aqueous solutions. The electrode showed a stable response towards H2O2after electrochemical activation. Effects of flow rate, operating potential, concentration, injection volume and interferences were investigated. A linear response towards H2O2from 5 μg.l−1to 450 mg.l−1and a detection limit (3 signal-to-noise ratio) of 4.7 μg.l−1was found. The method was employed for the determination of H2O2in rain water samples.

 