首页   按字顺浏览 期刊浏览 卷期浏览 Oxford Optronix MPM 3S: a clinical assessment of a microvascular perfusion monitor
Oxford Optronix MPM 3S: a clinical assessment of a microvascular perfusion monitor

 

作者: DrydenC. M.,   GrayW. M.,   AsburyA. J.,  

 

期刊: Journal of Medical Engineering&Technology  (Taylor Available online 1992)
卷期: Volume 16, issue 4  

页码: 159-162

 

ISSN:0309-1902

 

年代: 1992

 

DOI:10.3109/03091909209030220

 

出版商: Taylor&Francis

 

数据来源: Taylor

 

摘要:

The Oxford Optronix MPM 3S is a new microvascular perfusion monitor which is promoted as a device for use in the operating theatre. It uses a semiconductor laser diode and applies the Doppler principle to derive a semi-quantitative estimation of microvascular flow. We assessed this instrument with eight healthy volunteers who each performed eight different orlhostalic arm manoeuvres while forearm skin blood flow was monitored. The different manoeuvres caused statistically significant changes in the instrument's reading which generally were consistent with expected changes in blood flow. The monitor also was assessed in the theatre environment with four anaesthetized patients. It proved easy to use, and was not subject to electrical interference from other equipment including short-wave diathermy. The major practical limitation of the technique is the semi-quantitative nature of the measurement. The instrument appears to have potential clinical uses in plastic and vascular surgery.

 

点击下载:  PDF (490KB)



返 回