High‐accuracy optical attenuator continuously variable in the range of 1:20
作者:
Mikio Yamashita,
Yasushi Kawada,
Satoshi Takeuchi,
期刊:
Review of Scientific Instruments
(AIP Available online 1985)
卷期:
Volume 56,
issue 3
页码: 478-479
ISSN:0034-6748
年代: 1985
DOI:10.1063/1.1138272
出版商: AIP
数据来源: AIP
摘要:
A continuously variable optical attenuator is described that uses optical fiber light guides and high‐accuracy dial micrometers mounted on mechanical stages. The attenuation factors thereby obtained have proved reproducible to ∼0.1% over a dynamic range of 1:20. It should, therefore, be particularly useful in testing the linearity of the response of various photodetectors.
点击下载:
PDF
(167KB)
返 回