首页   按字顺浏览 期刊浏览 卷期浏览 Serum Thyroxine Binding Capacity-Dependent Bias in an Automated Free Thyroxine Assay
Serum Thyroxine Binding Capacity-Dependent Bias in an Automated Free Thyroxine Assay

 

作者: N.D. Christofides,   E. Wilkinson,   M. Stoddart,   D.C. Ray,   G.J. Beckett,  

 

期刊: Journal of Immunoassay  (Taylor Available online 1999)
卷期: Volume 20, issue 4  

页码: 201-221

 

ISSN:0197-1522

 

年代: 1999

 

DOI:10.1080/01971529909349351

 

出版商: Taylor & Francis Group

 

关键词: AXSYM Free Thyroxine;Binding capacity;Equilibrium dialysis;Protein-bound T4;immunoassay

 

数据来源: Taylor

 

摘要:

The magnitude of serum thyroxine (T4) binding capacity (sBC) dependent bias in the AXSYM free thyroxine (FT4) assay was assessed using two recently described tests. One of the tests uses a direct equilibrium dialysis (ED) FT4 assay as the reference method. the results obtained with the AXSYM method were compared with those obtained by the ED FT4 method in patient sera having a wide range of sBC. the other test involves comparison of the FT4 results obtained following dilution of sera by an inert buffer, to theoretically derived FT4 results. As serum dilution causes a predictable decrease in sBC, the demonstration of a negative bias whose magnitude increases in parallel to the dilution, is indicative of an sBC-dependent bias. the AXSYM FT4 assay exhibited a significant sBC-dependent bias. This sBC-dependent bias is likely to have been caused by the presence of significant amounts of T4 binding proteins in the assay reagents.

 

点击下载:  PDF (716KB)



返 回