首页   按字顺浏览 期刊浏览 卷期浏览 Effects of x‐ray tube current and voltage on effective focalspot size
Effects of x‐ray tube current and voltage on effective focalspot size

 

作者: Edward L. Chaney,   William R. Hendee,  

 

期刊: Medical Physics  (WILEY Available online 2003)
卷期: Volume 1, issue 3  

页码: 141-147

 

ISSN:0094-2405

 

年代: 2003

 

DOI:10.1118/1.1637286

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

In general, the dimensions of the effective focal spot of an x‐ray tube vary with tube current and voltage. This dependence on tube operating conditions has been noted in the literature but has not been investigated in detail. Star phantom measurements of seven focal spots for four x‐ray tubes operated at a variety of tube voltages and currents show that effective focal‐spot size varies asIV − 3 / 2. This dependence may be suppressed for the focal‐spot dimension perpendicular to the tube axis, however, because of auxiliary electrostatic focusing. Experimental data are presented with a theoretical explanation of the relationship between effective focal‐spot size and tube operating conditions.

 

点击下载:  PDF (503KB)



返 回