首页   按字顺浏览 期刊浏览 卷期浏览 The effect of using silicon based diffusion pump fluid on spectral quality in an electr...
The effect of using silicon based diffusion pump fluid on spectral quality in an electrospray ionization ion trap/time-of-flight mass spectrometer

 

作者: Randy W. Purves,   Wojciech Gabryelski,   Liang Li,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1997)
卷期: Volume 68, issue 8  

页码: 3252-3253

 

ISSN:0034-6748

 

年代: 1997

 

DOI:10.1063/1.1148276

 

出版商: AIP

 

数据来源: AIP

 

摘要:

The formation of adducts resulting from interactions between protein ions and residual gases in an electrospray ionization ion trap/time-of-flight mass spectrometer is reported. This instrument utilizes diffusion pumps to achieve high vacuum, and we found that when silicon based diffusion pump fluids were used, characteristic adduct peaks appeared in the mass spectra. These adduct peaks were not detected, however, when a polyphenyl ether (Santovac 5) was used as the diffusion pump fluid. In switching the diffusion pump fluid to Santovac 5, a rigorous cleaning procedure capable of removing residual silicon pump fluid from the instrument was a critical step. This cleaning procedure can also be used for routine cleaning of the ion trap and other source components. ©1997 American Institute of Physics.

 