Clinical Comparison of Ocular Blood Flow Tonometer, Tonopen, and Goldmann Applanation Tonometer for Measuring Intraocular Pressure in Postkeratoplasty Eyes

 

作者: Veena Rao,   Lawrence Gnanaraj,   Keith Mitchell,   Francisco Figueiredo,  

 

期刊: Cornea  (OVID Available online 2001)
卷期: Volume 20, issue 8  

页码: 834-838

 

ISSN:0277-3740

 

年代: 2001

 

出版商: OVID

 

关键词: Intraocular pressure;Goldmann applanation tonometer;Tonopen;Ocular blood flow tonometer;Penetrating keratoplasty;Pachymetry

 

数据来源: OVID

 

摘要:

Purpose.Intraocular pressure (IOP) measurements vary with different tonometers in irregular corneas. The purpose of our study was to determine the comparability of the ocular blood flow tonometer (OBF) and Tonopen with the Goldmann applanation tonometer (GAT) in measuring IOP in post-keratoplasty eyes.Methods.The IOP was measured using the OBF tonometer, GAT, and Tonopen in 69 post-keratoplasty eyes by two observers. The same instruments were used, and the sequence of measurements was randomized. Pachymetry to measure corneal thickness was done in all patients.Results.Interobserver variation for IOP measured with GAT was statistically significant (p= 0.039). Two-way analysis of variance (ANOVA), however, did not demonstrate any statistical difference between measurements with the three tonometers and two observers. Corneal thickness did not appear to have an effect on measurements with any of the tonometers.Conclusion.There was no significant difference in the overall accuracy of the OBF tonometer over the Tonopen versus the GAT. The OBF tonometer can be a useful alternative in measuring IOP in post-keratoplasty eyes.

 

点击下载:  PDF (4294KB)



返 回