首页   按字顺浏览 期刊浏览 卷期浏览 Comparisons of Methods for Calculating Retention and Separation of Chromatographic Peaks
Comparisons of Methods for Calculating Retention and Separation of Chromatographic Peaks

 

作者: R.E. Pauls,   L.B. Rogers,  

 

期刊: Separation Science  (Taylor Available online 1977)
卷期: Volume 12, issue 4  

页码: 395-413

 

ISSN:0037-2366

 

年代: 1977

 

DOI:10.1080/00372367708058085

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

The accuracy and precision of calculating retention times from means and peak maxima have been examined using an exponentially modified Gaussian as a model for tailed chromatographic peaks. At different levels of random noise, retention times could be determined with nearly the same precision using either the mean or maximum. However, the accuracies and precisions of the maxima were affected by the number of points used in the digital smooth and by the number of points recorded per unit of standard deviation. For two peaks of similar shape, consistency in the selection of points should usually permit differences in retention to be determined accurately and with approximately the same precision using maxima, means, or half-heights on the leading side of the peak.

 

点击下载:  PDF (672KB)



返 回