首页   按字顺浏览 期刊浏览 卷期浏览 Errors in Luminescence Decay Time Measurements Caused by Photomultiplier Saturation
Errors in Luminescence Decay Time Measurements Caused by Photomultiplier Saturation

 

作者: P. F. Jones,   A. R. Calloway,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1973)
卷期: Volume 44, issue 9  

页码: 1393-1395

 

ISSN:0034-6748

 

年代: 1973

 

DOI:10.1063/1.1686393

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A possible source of error in measuring luminescent decay times as a result of photomultiplier saturation is discussed. Photomultiplier saturation can occur by exceeding the maximum linear output current, which decreases with increased light pulse duration. For this reason it is necessary to calibrate the instrument response appropriate to the lifetime and light intensity being measured.

 