首页   按字顺浏览 期刊浏览 卷期浏览 Suppression of Demand Pacemakers by Inactive Pacemaker Electrodes
Suppression of Demand Pacemakers by Inactive Pacemaker Electrodes

 

作者: Warren Widmann,   Stelio Mangiola,   Lawrence Lubow,   Fayetta Dolan,  

 

期刊: Circulation  (OVID Available online 1972)
卷期: Volume 45, issue 2  

页码: 319-323

 

ISSN:0009-7322

 

年代: 1972

 

出版商: OVID

 

关键词: Electrode interference;Ventricular inhibited pacemaker;Pacemaker failure;Pacemaker inhibition

 

数据来源: OVID

 

摘要:

Ventricular inhibited cardiac pacemakers were suppressed by intimate contact of the tip of the active pacemaker electrode with that of the inactive pacemaker electrode in two patients. This contact apparently was sensed as myocardial electrical activity and resulted in variable suppression of pacemaker emission. There was no interference with R-wave sensing function. The threshold for stimulation was unaffected. There was no loss of capture as tested with the magnet-controlled continuous mode of stimulation. This type of pacemaker failure can be remedied simply by withdrawing the tips of the electrodes from intimate contact.

 