首页   按字顺浏览 期刊浏览 卷期浏览 Extended Mass Range Operation of the Monopole Mass Spectrometer
Extended Mass Range Operation of the Monopole Mass Spectrometer

 

作者: R. E. Grande,   R. L. Watters,   J. B. Hudson,  

 

期刊: Journal of Vacuum Science and Technology  (AIP Available online 1966)
卷期: Volume 3, issue 6  

页码: 329-333

 

ISSN:0022-5355

 

年代: 1966

 

DOI:10.1116/1.1492497

 

出版商: American Vacuum Society

 

数据来源: AIP

 

摘要:

Data are presented demonstrating the degree to which the mass range of monopole and quadrupole mass spectrometers may be extended by various techniques and concomittant effects that this mass range extension has on the other operating parameters such as resolving power, sensitivity, and mass discrimination. These data indicate that high sensitivity and unit resolution may be maintained over a mass range of greater than 600 amu in a monopole instrument of modest size. Further extension is also possible if resolution or sensitivity may be sacrificed.

 

点击下载:  PDF (446KB)



返 回