Linear accelerator calibration monitor with a memory

 

作者: Robert L. Dixon,   Kenneth E. Ekstrand,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 6, issue 5  

页码: 436-440

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.594533

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

A calibration monitor has been designed for measuring the constancy of linear accelerator or cobalt unit output between full calibrations. This monitor is battery‐operated, light‐weight, and slides into the shadow tray attachment on a linear accelerator or cobalt unit for easy setup. It provides a digital readout of the dose delivered, and a consistency check can be made in less than two minutes. The precision of the monitor, determined by cobalt‐60 irradiations over a 2 1/2 month period, is±0.6% (standard deviation). The monitor also retains the dose reading in a CMOS digital counter indefinitely, hence it can be used in the same fashion as mailed thermoluminescent dosimeters (TLD) for calibration checks at remote facilities without the complicated readout procedures associated with TLD. The monitor can be mailed to a remote facility, positioned without ambiguity, and irradiated; and the reading can be verified on return to the originating center simply by pressing a switch. The monitor can easily be set up to carry out a ’'blind” check in which the reading obtained is not known to the remote facility.

 