Emergency PaO2estimates in one minute with a transcutaneous oxygen sensor

 

作者: ROBERT,   APPLEBAUM RICHARD,   BOWMAN KEVIN,   TREMPER WILLIAM,  

 

期刊: Critical Care Medicine  (OVID Available online 1981)
卷期: Volume 9, issue 10  

页码: 742-743

 

ISSN:0090-3493

 

年代: 1981

 

出版商: OVID

 

数据来源: OVID

 

摘要:

Blood was applied directly to the electrode surface of a transcutaneous oxygen sensor (PtcO2) and the oxygen tension value obtained was compared to oxygen tension value from a conventional blood gas machine. Three hundred and seventeen blood samples were analyzed at four PtcO2electrode temperatures: 37, 40, 42, and 45°C. A linear regression of PO2vs. PtcO2blood drop (PbdO2) produced correlation coefficients (r) from 0.99 to 0.93 and standard errors from 3.2 to 4.7 torr at these temperatures.The mean time for stabilization was 51 sec. The PO2ranged from 4–150 torr. The authors conclude that an accurate PO2of blood can be obtained within 1 min by placing a drop of blood on the surface of a PtcO2electrode.

 