作者: D. A. Goloskie,   T. B. Stoughton,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1983)
卷期: Volume 54, issue 2  

页码: 177-182

 

ISSN:0034-6748

 

年代: 1983

 

DOI:10.1063/1.1137323

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A low‐pressure, high‐flow, recirculating gas system has been designed and constructed for use on a large volume (250 ft3) ionization sampling drift chamber (CRISIS), which operates at an absolute pressure of less than 1 in. of water relative to atmospheric pressure. The system was also designed to control and monitor the detector gas mixture and to reduce the oxygen level in the detector below 4 ppm. In operation the system demonstrated very stable long and short term stability and also demonstrated the gas cost savings inherent in recirculating systems.

 