首页   按字顺浏览 期刊浏览 卷期浏览 Fabrication of a238Pu Heat Source for a Cardiac Pacemaker
Fabrication of a238Pu Heat Source for a Cardiac Pacemaker

 

作者: FlemingD. L.,   BrewerC. O.,   CoffeyD. L.,  

 

期刊: Nuclear Applications and Technology  (Taylor Available online 1970)
卷期: Volume 8, issue 6  

页码: 536-538

 

ISSN:0550-3043

 

年代: 1970

 

DOI:10.13182/NT70-A28654

 

出版商: Taylor&Francis

 

数据来源: Taylor

 

摘要:

A238Pu-Powered heat source for use in an experimental model of a cardiac pacemaker was developed. The encapsulated source has a thermal power of 0.134±0.013 W and is fueled with three238Pu metal pins. Each fueled pin was TIC welded into a tantalum inner capsule. The inner capsules were electron beam welded into Hastelloy-C middle capsules. These middle capsules were platinum plated to prevent a reaction between the Hastelloy-C and the titanium outer capsules. Three middle capsules were electron beam welded into each outer capsule.

 