首页   按字顺浏览 期刊浏览 卷期浏览 DIFFERENTIAL EQUATION FOR THE USE OF THE DIFFUSED RING PHOTOKERATOSCOPE
DIFFERENTIAL EQUATION FOR THE USE OF THE DIFFUSED RING PHOTOKERATOSCOPE

 

作者: Sami El Hage,  

 

期刊: American Journal of Optometry and Archives of American Academy of Optometry  (OVID Available online 1972)
卷期: Volume 49, issue 5  

页码: 422-435

 

ISSN:0002-9408

 

年代: 1972

 

出版商: OVID

 

数据来源: OVID

 

摘要:

When a small diaphragm is placed at the back focal plane of the photokeratoscope, a differential equation which can determine the corneal topology in one plane is obtained. The solution of the differential equation results in a more accurate description of corneal topology than previous methods. The solution of the equation requires computer programs.For each keratoscope ring of a variable coordinate a and a fixed coordinate b, a differential equation is obtained. One differential equation characteristic of the corneal surface was sought. For the discret values are the polynomial function a1= a(y) was substituted. The numerical precision was demonstrated to be exact enough for practical purposes. Experimental errors had a minor influence on keralogram evaluation. Results were corrected for corneal misalignment. Corrections for defocusing require very precise measurement of the smallest ring image diameter. When the corneal vertex-position was aligned with an accuracy of 0.2 mm, the accuracy of measurement at the corneal periphery was better than 0.02 mm.

 

点击下载:  PDF (443KB)