Pulse radiolysis equipment: A setup for simultaneous multiwavelength kinetic spectroscopy
作者:
Manfred Saran,
Georg Vetter,
Michael Erben‐Russ,
Robert Winter,
Alf Kruse,
Christa Michel,
Wolf Bors,
期刊:
Review of Scientific Instruments
(AIP Available online 1987)
卷期:
Volume 58,
issue 3
页码: 363-368
ISSN:0034-6748
年代: 1987
DOI:10.1063/1.1139289
出版商: AIP
数据来源: AIP
摘要:
A setup for pulse radiolysis experiments is described consisting of the following main components: an array of 15 photomultipliers attached to a spectrograph (allowing 4032 spectra to be recorded with a time resolution of 500 ns per spectrum), self‐regulating high‐voltage supply for the photomultipliers, computer‐controlled solution mixing and dispensing system, and provisions for continuous dose variation of the FEBETRON‐accelerator electron beam. The general performance of the system is discussed and construction or electronic details are given for special components.
点击下载:
PDF
(667KB)
返 回