Determination and optimization of frequency response of constant temperature hot‐wire anemometers in supersonic flows
作者:
J. P. Bonnet,
T. Alziary de Roquefort,
期刊:
Review of Scientific Instruments
(AIP Available online 1980)
卷期:
Volume 51,
issue 2
页码: 234-239
ISSN:0034-6748
年代: 1980
DOI:10.1063/1.1136180
出版商: AIP
数据来源: AIP
摘要:
In order to achieve turbulence measurements in supersonic flows the frequency response of CTA is tested and improved. A fast and convenient technique to determine electrically the frequency response is proposed; the response is also tested by direct heating of the wire by means of a laser beam. The two techniques are in good agreement. Under certain experimental conditions (M=2,Re&fgr;=400) it is possible to use CTA with a satisfactory frequency response up to 250 kHz and overheat ratios down to 0.07. In these severe operating conditions, the square‐wave test technique seems to be not sufficient to ensure optimum settings and it is necessary to use symmetrical bridges with fine adjustment of offset voltage.
点击下载:
PDF
(446KB)
返 回