首页   按字顺浏览 期刊浏览 卷期浏览 Automatic Temperature Control for an Adiabatic Calorimeter
Automatic Temperature Control for an Adiabatic Calorimeter

 

作者: W. W. Cleland,   Richard S. Harding,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1957)
卷期: Volume 28, issue 9  

页码: 696-698

 

ISSN:0034-6748

 

年代: 1957

 

DOI:10.1063/1.1715979

 

出版商: AIP

 

数据来源: AIP

 

摘要:

An automatic temperature control device for an adiabatic oxygen bomb calorimeter has been developed which keeps the temperature of the outer water jacket equal to that of the inner bath. When the temperature of the inner bath is not changing rapidly, heat losses are negligible; however, the error due to lag when the inner bath temperature is rising rapidly is about two parts per thousand. This degree of precision permits the accurate analysis of samples of low caloric content such as lyophilized urine. Since the technician need not watch the calorimeter once the bomb has been fired, but instead can prepare a second bomb, the number of samples analyzed in a given time can be doubled if two bombs are available for use with the calorimeter.

 

点击下载:  PDF (237KB)



返 回