首页   按字顺浏览 期刊浏览 卷期浏览 Cryostat for annealing spectra measurements with short rise time to the annealing tempe...
Cryostat for annealing spectra measurements with short rise time to the annealing temperature

 

作者: P. Andronikos,   E. Olympios,   C. Papathanassopoulos,  

 

期刊: Radiation Effects  (Taylor Available online 1971)
卷期: Volume 10, issue 4  

页码: 215-218

 

ISSN:0033-7579

 

年代: 1971

 

DOI:10.1080/00337577108231088

 

出版商: Taylor & Francis Group

 

数据来源: Taylor

 

摘要:

A simple cryostat for annealing spectra measurements in the temperature range 78 to 700 °K is described and its specifications are given. The rise time at the annealing temperature is chosen equal to 1.5 minutes, giving thus, a practically square shape of the temperature pulse. An analogue valve regulates automatically a stream of cold nitrogen gas and maintains the temperature of the samples constant within ±0.05 °C. The rise time varies from 1 min, or even less, to about 7 min, when there is no control of temperature through the cold gas.

 

点击下载:  PDF (301KB)



返 回