首页   按字顺浏览 期刊浏览 卷期浏览 A simple and effective method for improving the accuracy of selective audio‐freq...
A simple and effective method for improving the accuracy of selective audio‐frequency voltage measurements

 

作者: Obradovic Mirko,   Rodic Ranko,   Arandjelovic Vojislav,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 5  

页码: 1711-1713

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1145212

 

出版商: AIP

 

数据来源: AIP

 

摘要:

An examination of the results of the accuracy measurements of a bench top audio‐frequency selective voltmeter, based on the superheterodyne principle, showed that an accuracy of the order of only 10% of measured values could be achieved if no internal error correction procedure in the measurements had been used. The temperature variations of the parameters of the quartz band‐pass filter have been identified as the main source of the measurement error. An application of the internal reference signal from a relatively simple and inexpensive source improved the overall accuracy up to the order of 3%. Finally, it was shown that by combining the wide band and selective measurements of the signals from the built‐in tracking generator further improved the overall accuracy to better than 1% and eliminated altogether the need for using any separate source of internal reference signals.

 