首页   按字顺浏览 期刊浏览 卷期浏览 Time-Resolved Spark-Source Mass Spectrometry; The Effect Of Spark Duration On Ion Inten...
Time-Resolved Spark-Source Mass Spectrometry; The Effect Of Spark Duration On Ion Intensity And Precision Of Analysis

 

作者: JamesC. Franklin,   JohnA. Dean,  

 

期刊: Analytical Letters  (Taylor Available online 1971)
卷期: Volume 4, issue 12  

页码: 857-865

 

ISSN:0003-2719

 

年代: 1971

 

DOI:10.1080/00032717108066072

 

出版商: Taylor & Francis Group

 

关键词: mass spectrometry;spark source;analysis;time resolved;ion intensity;precision

 

数据来源: Taylor

 

摘要:

A radio-frequency probe and a synchronized electronic ion-beam chapper have been used to determine ion-intensity variation with spark duration. Large peaks in multiply charged ion intensities are detected in the early stages of the radio-frequency spark, while later in the spark period the singly charged species predominate. The effects of these variations on the precision of analysis have been determined. A gate was selected in the spark pulse, and a factor of two increase in precision was obtained.

 

点击下载:  PDF (313KB)



返 回