首页   按字顺浏览 期刊浏览 卷期浏览 Effect of half‐wavelength membranes on the axial resolution of real‐time ultrasonic sca...
Effect of half‐wavelength membranes on the axial resolution of real‐time ultrasonic scanners

 

作者: Jonathan Ophir,   P. A. Narayana,  

 

期刊: The Journal of the Acoustical Society of America  (AIP Available online 1983)
卷期: Volume 74, issue 3  

页码: 687-690

 

ISSN:0001-4966

 

年代: 1983

 

DOI:10.1121/1.389853

 

出版商: Acoustical Society of America

 

数据来源: AIP

 

摘要:

The effect of interposing a half‐wavelength membrane between the transducer and the load in mechanically sectored ultrasonic scanners was investigated. In particular, the effects of impedance mismatch and detuning on broadband spectra were theoretically and experimentally studied. These studies indicate that for certain common cases, a bandwidth reduction of 20%–30% can be anticipated, resulting in a proportional decrease in the axial resolution.

 