首页   按字顺浏览 期刊浏览 卷期浏览 Signal-to-noise improvement of bio-magnetic signals using a flux-gate probe and real ti...
Signal-to-noise improvement of bio-magnetic signals using a flux-gate probe and real time signal processing

 

作者: C. Dolabdjian,   S. Saez,   A. Reyes Toledo,   D. Robbes,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1998)
卷期: Volume 69, issue 10  

页码: 3678-3680

 

ISSN:0034-6748

 

年代: 1998

 

DOI:10.1063/1.1149158

 

出版商: AIP

 

数据来源: AIP

 

摘要:

We have recorded the magnetic signal generated by the human cardiac muscle with the help of a very sensitive flux-gate sensor. First, we have verified the feasibility of such detection inside a magnetically shielded room, then repeated this same experiment within our laboratory and in a low shielded environment. Results have shown the easy operation of a flux-gate sensor for this purpose and the efficiency of fast numerical processing in a low shielded environment. This leads in both cases to a voltage signal-to-noise ratio of about 5 within a 100 Hz bandwidth, which corresponds to the actual limit of the flux-gate probe.©1998 American Institute of Physics.

 

点击下载:  PDF (66KB)



返 回