Effect of Cardiac Output on Extravascular Lung Water Measurements Made with the Thermoconductivity Method
作者:
William,
Noble J.,
期刊:
Anesthesiology
(OVID Available online 1986)
卷期:
Volume 65,
issue 2
页码: 206-209
ISSN:0003-3022
年代: 1986
出版商: OVID
关键词: Anesthetics, volatile: halothane.;Heart: cardiac output.;Lung: extravascular lung water.;Measurement techniques: cardiac output; extravascular lung water.;Sympathetic nervous system sympathomimetic agents: isoproterenol.
数据来源: OVID
摘要:
The authors tested the effects of injectate temperature and of changing cardiac output on the measurement of lung water by the thermoconductivity technique (ETVL). Cardiac output in dogs was increased and decreased with isoproterenol and halothane, respectively. Post mortem extravascular lung water (pulmonary extravascular tissue weight [PETW]) was determined using a weighing technique. Cardiac output varied between 0.7 and 8.8 1/min and did not influence the ETVLmeasurement. The authors conclude ETVLmeasured by the thermoconductivity technique is not influenced by large changes in cardiac output.