Accuracy of automatic tube compensation in new-generation mechanical ventilators*

 

作者: Serge,   Elsasser Josef,   Guttmann Reto,   Stocker Georg,   Mols Hans-Joachim,   Priebe Christoph,  

 

期刊: Critical Care Medicine  (OVID Available online 2003)
卷期: Volume 31, issue 11  

页码: 2619-2626

 

ISSN:0090-3493

 

年代: 2003

 

出版商: OVID

 

关键词: endotracheal tube resistance;tube compensation;work of breathing;additional work of breathing;expiratory work of breathing

 

数据来源: OVID

 

摘要:

ObjectiveTo compare performance of flow-adapted compensation of endotracheal tube resistance (automatic tube compensation, ATC) between the original ATC system and ATC systems incorporated in commercially available ventilators.DesignBench study.SettingUniversity research laboratory.SubjectsThe original ATC system, Dräger Evita 2 prototype, Dräger Evita 4, Puritan-Bennett 840.InterventionsThe four ventilators under investigation were alternatively connected via different sized endotracheal tubes and an artificial trachea to an active lung model. Test conditions consisted of two ventilatory modes (ATC vs. continuous positive airway pressure), three different sized endotracheal tubes (inner diameter 7.0, 8.0, and 9.0 mm), two ventilatory rates (15/min and 30/min), and four levels of positive end-expiratory pressure (0, 5, 10, and 15 cm H2O).Measurements and Main ResultsPerformance of tube compensation was assessed by the amount of tube-related (additional) work of breathing (WOBadd), which was calculated on the basis of pressure gradient across the endotracheal tube. Compared with continuous positive airway pressure, ATC reduced inspiratory WOBaddby 58%, 68%, 50%, and 97% when using the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. Depending on endotracheal tube diameter and ventilatory pattern, inspiratory WOBaddwas 0.12–5.2 J/L with the original ATC system, 1.5–28.9 J/L with the Puritan-Bennett 840, 10.4–21.0 J/L with the Evita 2 prototype, and 10.1–36.1 J/L with the Evita 4 (difference between each ventilator at identical test situations,p< .025). Expiratory WOBaddwas reduced by 5%, 26%, 1%, and 70% with the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. The expiratory WOBaddcaused by an endotracheal tube of 7.0 mm inner diameter was 5.5–42.2 J/L at a low ventilatory rate and 19.6–82.3 J/L at a high ventilatory rate. It was lowest with the original ATC system and highest with the Evita 4 ventilator (p< .025).ConclusionsFlow-adapted tube compensation by the original ATC system significantly reduced tube-related inspiratory and expiratory work of breathing. The commercially available ATC modes investigated here may be adequate for inspiratory but probably not for expiratory tube compensation.

 

点击下载:  PDF (571KB)



返 回