Evaluation of a replenishment type oxygen consumption monitor
作者:
DWAYNE WESTENSKOW,
STEVE ROBERTS,
NATHAN PACE,
期刊:
Critical Care Medicine
(OVID Available online 1989)
卷期:
Volume 17,
issue 8
页码: 798-802
ISSN:0090-3493
年代: 1989
出版商: OVID
数据来源: OVID
摘要:
This paper describes the testing of a monitor (Oxyconsumeter) which uses a replenishment technique to measure oxygen consumption (VO2). Oxygen is added to the expired gas to replace the oxygen consumed by the patient. The replenishing oxygen flow equals the patient's &OV0312;O2when the mixed replenished expired oxygen fraction equals the inspired oxygen fraction. The accuracy of the Oxyconsumeter was assessed using a N2and CO2dilution technique. When tested over the operating range specified by the manufacturer, the bias of the Oxyconsumeter was 1.66% of reading. The reproducibility averaged 6.96% of reading. The reproducibility did not change with &OV0312;O2, FIO2or minute ventilation. The replenishment technique has the theoretical advantage that an absolute oxygen sensor calibration is not necessary and a respiratory flowmeter is not needed. It appears, from the laboratory testing, that the Oxyconsumeter has sufficient accuracy for use in the critical care setting.