首页   按字顺浏览 期刊浏览 卷期浏览 An Optical Fluorescence System for Quantitative Pressure Measurement in the Diamond&hyp...
An Optical Fluorescence System for Quantitative Pressure Measurement in the Diamond‐Anvil Cell

 

作者: J. D. Barnett,   S. Block,   G. J. Piermarini,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1973)
卷期: Volume 44, issue 1  

页码: 1-9

 

ISSN:0034-6748

 

年代: 1973

 

DOI:10.1063/1.1685943

 

出版商: AIP

 

数据来源: AIP

 

摘要:

An optical system for rapid routine pressure measurement is described which utilizes a pressure shift in the sharpR‐line fluorescence spectrum of ruby or similar materials. The system, which consists of a standard polarizing microscope and a 1/4 m monochromator with associated photodetection system, is used with the diamond‐anvil pressure cell, but can be employed with suitable modification in any pressure system which has optical access. The precision of the pressure measurement in a hydrostatic environment up to 100 kilobar is 0.5 kilobar using ruby as the pressure sensor. This precision is better than the accuracy of the present pressure scale above 40 kilobar. The merits of fluorescent materials other than ruby as pressure sensors are also discussed. A description of a Waspaloy diamond cell with some modifications in design is given. This improved cell and associated techniques extends the pressure range in gasketed systems up to 200 kilobar at room temperature and to moderate pressures at 700° C.

 

点击下载:  PDF (725KB)