Laser‐heated high‐temperature EPR spectroscopy
作者:
T. A. Yager,
W. D. Kingery,
期刊:
Review of Scientific Instruments
(AIP Available online 1980)
卷期:
Volume 51,
issue 4
页码: 464-466
ISSN:0034-6748
年代: 1980
DOI:10.1063/1.1136246
出版商: AIP
数据来源: AIP
摘要:
A CO2laser‐heated sample holder for an EPR spectrometer capable of temperatures above 1200°C is described. Since only the sample is heated, this system permits rapid temperature changes and rotation of the magnet around the sample. The spectra of MgO containing Fe3+and Cr3+at room temperature and 1200°C are presented.