Soft x‐ray tomography diagnostic at the RTP tokamak

 

作者: D. F. da Cruz,   A. J. H. Donne´,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 7  

页码: 2295-2305

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144680

 

出版商: AIP

 

数据来源: AIP

 

摘要:

An 80‐channel soft x‐ray tomography system has been constructed for diagnosing the RTP (Rijnhuizen Tokamak Project) tokamak plasma. Five pinhole cameras, each with arrays of 16 detectors are distributed more or less homogeneously around a poloidal plasma cross section. The cameras are positioned close to the plasma edge to enable viewing angles which could not be reached otherwise. The system can be used for making tomographic reconstructions of the soft x‐ray emissivity with a relatively high poloidal mode number (up to both components ofm=4), which is an improvement with respect to other plasma tomography systems presented in literature. The temporal resolution of the system is 5 &mgr;s. Apart from a description of the hardware and geometrical layout of the system, the paper will focus on two important issues, namely the calibration of the detectors and the positioning of the cameras in the vacuum vessel of the tokamak. A dedicated x‐ray source has been designed for the relative calibration of the detectors of each individual camera. Misalignment causes several artifacts in tomographic reconstructions obtained with multicamera systems. Therefore, two new methods are proposed which make use of the plasma itself for improving the knowledge about the positions of the cameras.  

 