首页   按字顺浏览 期刊浏览 卷期浏览 In VitroStress Analyses of Dental Implants Supporting Screw‐Retained and Cement&...
In VitroStress Analyses of Dental Implants Supporting Screw‐Retained and Cement‐Retained Prostheses

 

作者: Wook,   Kim Zhimon,   Jacobson Dan,  

 

期刊: Implant Dentistry  (OVID Available online 1999)
卷期: Volume 8, issue 2  

页码: 141-151

 

ISSN:1056-6163

 

年代: 1999

 

出版商: OVID

 

关键词: provisional-cement-retained prosthesis;photoelastic analysis;strain gauge analysis;splinted implants;cantilever prostheses;single tooth implants;stress distribution

 

数据来源: OVID

 

摘要:

A photoelastic and strain gauge analysis was performed to evaluate the stress transferred to implants through the provisionalcement-retained, the permanentcement-retained, and the screwretained prostheses. The deflections of the prostheses at the time of the loading were also measured. In the single crown test, the provisionalcement-retained crowns transferred less stress. In the two-unit fixed partial denture test, there were no differences between the three different prostheses. In the two-implant supported distal cantilevered prostheses, the screw-type and the permanent-cement-retained prostheses developed more stress around the apex of both implants. The permanent-cement-retained prostheses acted almost the same as the screw-type.

 