Remote Control Systems' Radiotelemetry Receiver 7060: its use in clinical colonic motility measurement
作者:
ThorburnHeather A.,
CarterKenneth B.,
GoldbergJacqueline A.,
FinlayIan G.,
期刊:
Journal of Medical Engineering&Technology
(Taylor Available online 1989)
卷期:
Volume 13,
issue 5
页码: 252-256
ISSN:0309-1902
年代: 1989
DOI:10.3109/03091908909033164
出版商: Taylor&Francis
数据来源: Taylor
摘要:
A new radiotelemetry receiver type 7060 for use with radiopills is described. Its performance under laboratory conditions was assessed and its use for clinical colonic pressure measurements evaluated. The receiver, which resembles a small microcomputer, was easy to set up and worked reasonably well under ward conditions. Difficulties were experienced with the belt aerial supplied with the receiver, which did not reliably pick up signals from a freely moving radiopill. The receiver is not suitable for ambulatory applications. Its main clinical use is in situations where frequent sampling of the radiopill's signal over a long period of time is essential (e.g. colonic pressures). The large quantity of data generated can be fed from the receiver's analogue or digital output on line to a recorder.