首页   按字顺浏览 期刊浏览 卷期浏览 Audiometric Earphone Discomfort Level and Hearing Aid Saturation Sound Pressure Level f...
Audiometric Earphone Discomfort Level and Hearing Aid Saturation Sound Pressure Level for a 90 Decibel Input Signal (SSPL90) as Measured in the Human Ear Canal

 

作者: Arne Leijon,   Earl Hatford,   Gunnar Lidén,   Anders Ringdahl,   Ann-Katrin Dahlberg,  

 

期刊: Ear and Hearing  (OVID Available online 1983)
卷期: Volume 4, issue 4  

页码: 185-189

 

ISSN:0196-0202

 

年代: 1983

 

出版商: OVID

 

数据来源: OVID

 

摘要:

The acoustical problems involved in matching the saturation sound pressure level for a 90 dB input signal (SSPL90) of a hearing aid to individual discomfort level were investigated. The real ear SPL (RE/SSPL90) produced by a supra-aural earphone used when measuring uncomfortable loudness (UCL), and RE/SSPL90produced by three different hearing aids at 90 dB SPL input, were measured for nine subjects, using a miniature microphone technique, and compared to the corresponding coupler levels used when matching hearing aid maximum output to UCL. It was found that a hearing aid often gives about 5 dB, and sometimes 10 dB, higher RE/SPLs than the earphone, if the hearing aid output levels, as measured in a 2-cc coupler (IEC1261, are equal to the earphone output levels as measured in a 6-cc coupler (NBSSA). It is recommended that a safety margin of at least 5 dB be used in the preliminary fitting when matching hearing aid SSPL90to the patient's UCL, converted to dB SPL.

 

点击下载:  PDF (517KB)



返 回