首页   按字顺浏览 期刊浏览 卷期浏览 A continuous flow intermittent mandatory ventilation with continuous positive airway pr...
A continuous flow intermittent mandatory ventilation with continuous positive airway pressure circuit with high‐compliance reservoir bag

 

作者: ANTONIO BRASCHI,   GIORGIO IOTTI,   ALESSANDRO LOCATELLI,   GUIDO BELLINZONA,  

 

期刊: Critical Care Medicine  (OVID Available online 1987)
卷期: Volume 15, issue 10  

页码: 947-950

 

ISSN:0090-3493

 

年代: 1987

 

出版商: OVID

 

数据来源: OVID

 

摘要:

Unstable inspiratory pressure may reduce the benefits of intermittent mandatory ventilation and impose a greater work of breathing than some patients can tolerate. The authors evaluated a continuous flow intermittent mandatory ventilation with continuous positive airway pressure apparatus provided with a 10-L high compliance reservoir bag, separated from the ventilator circuit by a one-way valve. A pulmonary simulator was tested to generate a wide range of inspiratory peak flow rates. Pressures and flow rates were examined in different sections of the system, using PEEP levels up to 15 mm Hg and different fresh gas flow rates. The circuit offers many advantages, particularly considering the low inspiratory pressure drop and the low fresh gas flow rates needed for optimal function.

 

点击下载:  PDF (337KB)