Effect of catheter position on thermodilution cardiac output during continuous positive‐pressure ventilation

 

作者: FAYSAL HASAN,   ANTHONY MALANGA,   WILLIAM CORRAO,   SIDNEY BRAMAN,  

 

期刊: Critical Care Medicine  (OVID Available online 1984)
卷期: Volume 12, issue 4  

页码: 387-390

 

ISSN:0090-3493

 

年代: 1984

 

出版商: OVID

 

数据来源: OVID

 

摘要:

PEEP may decrease regional perfusion to nondependent lung regions, thereby creating different zones of thermal dissipation in dependent and nondependent zones of the lung. Under these conditions, the measurement of cardiac output by thermodilution may, thus, be influenced by the vertical position of the pulmonary artery catheter in the lung. We investigated this hypothesis in 7 healthy, anesthetized pigs by comparing cardiac output measurements from thermistors located in dependent and nondependent lung regions at varying levels of PEEP. Our data from thermistors in these 2 positions were similar, suggesting that the measurement of cardiac output by thermodilution is not influenced by the vertical position of the thermal sensor with respect to the left atrium.

 